• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SPACEOAR VUE SYSTEM; ABSORBABLE PERIRECTAL SPACER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SPACEOAR VUE SYSTEM; ABSORBABLE PERIRECTAL SPACER Back to Search Results
Model Number SV-2101
Device Problem Positioning Problem (3009)
Patient Problem Discomfort (2330)
Event Date 01/05/2024
Event Type  malfunction  
Manufacturer Narrative
Block d4 and h4: the complainant could not report the lot number; therefore, the manufacture and expiration dates are unknown.Block h6: imdrf device code a1502 captures the reportable event gel misplaced - non-vascular.
 
Event Description
It was reported to boston scientific corporation that a spaceoar vue device was implanted during a spaceoar vue placement procedure on (b)(6) 2024.The procedure was performed under monitored anesthesia care (mac) and fiducials were placed transperineally.During the procedure the needle was inadvertently placed into the prostate capsule and then repositioned.A computed tomography (ct) was performed and a foley catheter was placed.During simulation, it was noted the gel was partially in the correct location, and some hydrogel had entered the prostate capsule, and moved around the prostate.After the procedure, the patient experienced discomfort and the feeling of fullness.The patient will proceed with planned treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACEOAR VUE SYSTEM
Type of Device
ABSORBABLE PERIRECTAL SPACER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
500 commander shea boulevard
quincy MA 02171
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18664739
MDR Text Key334835450
Report Number2124215-2024-05078
Device Sequence Number1
Product Code OVB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182971
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSV-2101
Device Catalogue NumberSV-2101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/07/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-