It was reported that electrogram (egm) review was requested for this cardiac resynchronization therapy pacemaker (crt-p) system due ventricular fibrillation (vf) undersensing concerns.It was suspected there was inappropriate pacing into fine polymorphic ventricular tachycardia (pvt), as t-waves were visible when the device paced.Additionally, the patient passed out in the hospital.Upon review, technical services (ts) noted intermittent right ventricular (rv) undersensing.There was also a concern with the device clock as the patient passed out one hour apart from the stored episode, and it was discussed that the time discrepancy was likely attributed to daylight savings time.Right ventricular automatic threshold (rvat) was set to trend at 3v, and there was no evidence of loss of capture (loc).The highest threshold measurement was 2.3v and sensitivity was set to 2.5mv.Some stored pacemaker mediated tachycardia (pmt) episodes were noted, but appeared to be sinus tracking.This crt-p remains in service.No additional adverse patient effects were reported.
|