SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VH
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2024
Event Type  malfunction  
Event Description
It was reported, the cysto-nephro videoscope had reprocessing issues.The biomedical engineer (bme) at the facility emailed technical assistance center (tac) as he had concerned about scopes being improperly processed and transported.There were no reports of patient harm.
 
Manufacturer Narrative
The olympus endoscopic support specialist (ess) called and spoke to the biomedical engineer (bme) and provided the bme with reprocessing information from the instruction for use (ifu).The ess asked the bme is they would like onsite training to be provided but the bme declined and states he felt confident and had more understanding of the workflow of the cysto-nephro videoscope and the gi colon scope and transportation recommendation from the ifu's.As part of the investigation into this report, an olympus ess has been dispatched on-site to assess the reprocessing practices at the user facility.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation and correction to d4.The device history record was unable to be reviewed for this device.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the root cause of the insufficient/incorrect reprocessing of the scope was not determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18664806
MDR Text Key335417282
Report Number3002808148-2024-01202
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K221683
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VH
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/18/2024
Initial Date FDA Received02/07/2024
Supplement Dates Manufacturer Received03/11/2024
Supplement Dates FDA Received03/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-