Model Number 866199
Device Problems Device Alarm System (1012); Energy Output Problem (1431)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Event Description
This report is based on information provided by a philips remote service engineer (rse) and a philips field service engineer (fse) and has been investigated by the philips complaint handling team.Philips received a complaint on the fm100 defibrillator indicating that device issues the defibrillator disabled message, it is followed by a solid red-x on the rfu (ready for use) indicator.The event was outside of use and there was no reported patient nor user harm.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EFFICIA DFM100
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH 
Manufacturer (Section G)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH  
Manufacturer Contact
feng she
no.2 keji north 3rd road
nanshan district
shenzhen 
CH  
7552698099
MDR Report Key18665424
MDR Text Key334844818
Report Number3030677-2024-00511
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866199
Device Catalogue Number866199
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/19/2024
Initial Date FDA Received02/08/2024
Supplement Dates Manufacturer Received02/06/2024
Supplement Dates FDA Received02/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-