Catalog Number CS-25122-F |
Device Problems
Unraveled Material (1664); Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/02/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that (b)(6) 2023, the swg was found kinked during use on the patient.They were reported as fine post the procedure.
|
|
Manufacturer Narrative
|
(b)(4), the customer returned a single guide wire for evaluation.The guide wire was unraveled and showed evidence of use.Visual inspection of the guide wire revealed it was unraveled from the distal weld and was kinked in several locations along the body.The core wire distal j-bend was deformed and exposed out of the coil wire.Microscopic examination of the guide wire confirmed the core wire was broken adjacent to the distal weld.The exposed distal core wire tip is tapered and discolored at the point of separation.Both welds were present and appeared full and spherical.The major kinks in the guide wire body were measured at 293, 350, and 645 mm from the distal tip.The broken core wire measured 680 mm which is within the specifications of 678-688 mm per product drawing; this indicates that no pieces of the core wire appear to be missing.The outer diameter of the guide wire in an undamaged location measured 0.85 mm which is within the specification of 0.838-0.877 mm per product drawing.Functional inspection of the returned guide wire could not be performed due to the damage.A manual tug test confirmed that both the proximal weld was intact.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with the kit warns the user , "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished , radiographic visualization should be obtained and further consultation requested." the report that the guide wire unraveled was confirmed through examination of the returned sample.The core wire was broken adjacent to the distal weld.The guide wire and catheter met all dimensional requirements during investigation testing.No manufacturing defects were found during this investigation.Arrow guide wires of this size are designed and manufactured to withstand a tensile force of 2.75 pounds force.This internal specification is higher than the bs en iso 11070:2014 standard of 2.2 pounds force for this size wire.The selected insertion site and patient anatomy may present a tortuous path that could contribute to the possibility of guide wire kinking.Guide wire breakage may occur if a force greater than the design specification is applied during removal.Based on these circumstances, unintentional use error caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
|
|
Event Description
|
It was reported that 2 dec 2023, the swg was found kinked during use on the patient.They were reported as fine post the procedure.
|
|
Manufacturer Narrative
|
(b)(6) complaint verification testing could not be performed as no sample was returned for analysis.The instructions for use (ifu) provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained, and further consultation requested." a device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
|
Event Description
|
It was reported that"(b)(6) 2023, the swg was found kinked during use on the patient.They were reported as fine post the procedure.
|
|
Search Alerts/Recalls
|