• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number AFAPRO28
Device Problems Material Integrity Problem (2978); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2023
Event Type  malfunction  
Event Description
It was reported that during a cryo ablation procedure, a system notice was received indicating that the safety system detected a compromised outer vacuum.The balloon catheter was replaced which resolved the issue. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files and the afapro28 balloon catheter with lot number 20144 were returned and analyzed.The patient file showed seven applications were performed using the first balloon catheter identified as afapro28 balloon catheter of lot 20144.Additionally, the patient file showed ten applications were performed using the second balloon catheter identified as afapro28 balloon catheter of lot 20144.The patient file showed a system notice 50032(the safety system has detected a compromised outer vacuum) during ablation at application number seven.External visual inspection of the balloon segment showed blood/fluid inside the balloon.The catheter smart chip data was downloaded and reviewed.Data indicated the catheter was used for seven applications on the reported event date.During functional testing, the console terminated the application and triggered a system notice 50005(that the safety system detected fluid in the catheter and stopped the injection).During pressure testing of the balloon segment, a double-balloon breach condition was identified.Dissection did not show any signs of lifted thermocouple wires or leak detection wires that could have caused the breach.No anomalies were identified during inspection and/or pressure testing of the shaft segment.The pull wire, shaft, and guide wire lumen were intact.In conclusion, the balloon catheter failed the returned product inspection due to a double balloon breach.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18666565
MDR Text Key334840049
Report Number3002648230-2024-00042
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAFAPRO28
Device Catalogue NumberAFAPRO28
Device Lot Number20144
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/06/2024
Initial Date FDA Received02/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-