U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND IN.PACT 018 PACLITAXEL-COATED PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOO; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND IN.PACT 018 PACLITAXEL-COATED PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOO; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS Back to Search Results
Catalog Number IPU05012013P
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician was attempting to use an impact 018 drug coated balloon along with a 6fr non-medtronic sheath and a 300cm non-medtronic guidewire during treatment of a 100mm plaque lesion in the patients left distal superficial femoral artery (sfa).No vessel tortuosity and no vessel calcification were reported.Lesion exhibited cto (chronic complete occlusion: 100%) stenosis.Artery diameter is reported as 5mm.There were no anatomical abnormalities reported.Embolic protection was not used.There was no issue noted when removing the device from the hoop/tray.The device was prepped per ifu with no issues noted.The balloon inflated using an inflation device.Balloon twist was observed at 8atm.Even when expanded for 3 minutes the middle of the balloon would not be released.There were no abnormalities with the refill.No resistance occurred during delivery.The procedure was completed with this device.No balloon deflation difficulties were reported.The device was safely removed from patient.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN.PACT 018 PACLITAXEL-COATED PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOO
Type of Device
DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18667564
MDR Text Key334898621
Report Number9612164-2024-00657
Device Sequence Number1
Product Code ONU
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P140010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/08/2024
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIPU05012013P
Device Lot Number0011752691
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2024
Initial Date FDA Received02/08/2024
Date Device Manufactured04/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
-
-