|
Model Number 27761 |
Device Problems
Break (1069); Fluid/Blood Leak (1250); Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/07/2024 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding an external drainage and monitoring system.It was reported that the drain was opened and had a crack.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
The following regulatory reports are associated with this event: - 9612164-2024-00675 - 9612164-2024-00676 - 9612164-2024-00930 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Additional information received reported that the healthcare professional (hcp) went down to the emergency department to help with t he setup of the device with the emergency room staff per the request of the stroke coordinator due to the fact that they were attempting to set up the system and it was leaking from the priming of the system from the provider.The next three that were checked also had cracks near the same locations.The location on these devices were near where you hook-up the transducer to the stopcock.This would be the stopcock that led to the drainage system bag system-the bottom side of the stopcock.All three devices had hairline cracks through the tubing that you could feel with your finger.The patient did have a system placed with a different model.
|
|
Search Alerts/Recalls
|
|
|