• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. CLINITEK STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. CLINITEK STATUS+ Back to Search Results
Catalog Number 10379675
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2024
Event Type  malfunction  
Manufacturer Narrative
The customer stated that they no longer have any remaining reagent to return for investigation.They stated that they only perform maintenance if qc fails.They run qc once a week and it has been passing.Per the instrument manual, "clean the test table and table insert weekly or more frequently, if necessary, to maintain the analyzer for the following reasons: ensure that the analyzer operates properly, provide accurate test results, prevent contamination, avoid bacterial growth".The customer has since cleaned the calibration bar and is operational.A review of the certificate of analysis for the lot in use at the time of the event shows it was meeting specifications at the time of release.
 
Event Description
The customer reported that they received a false negative leukocyte result compared to retesting at another lab.It is unknown if the same sample was used for retesting.There is no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD.
northern road
chilton industrial estate
sudbury, suffolk CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
MDR Report Key18671398
MDR Text Key336301939
Report Number3002637618-2024-00007
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received02/08/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age24 MO
-
-