• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD LEGACY PCA PUMP; PUMP, INFUSION, PCA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD LEGACY PCA PUMP; PUMP, INFUSION, PCA Back to Search Results
Model Number 6300
Device Problem Inaccurate Flow Rate (1249)
Patient Problem Insufficient Information (4580)
Event Date 01/01/2024
Event Type  malfunction  
Event Description
It was reported that the device's delivery rate too low.There was unknown patient involvement.
 
Manufacturer Narrative
B3, g4, d4: udi unknown.(b)(6).H3: device not received by manufacturer.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Date returned to mfg (b)(6) 2024.One device was returned for evaluation.Visual inspection revealed the device in good condition.Event history log review was not applicable.Functional testing could not replicate the reported issue; no issues were found with the device delivery rate.The pump was found to be functioning properly and delivering within specification.The root cause was unknown.No further action was taken.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD LEGACY PCA PUMP
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan ln n
minneapolis, MN 55442
2247062300
MDR Report Key18671782
MDR Text Key334980875
Report Number3012307300-2024-00640
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6300
Device Catalogue Number21-6300-03
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/09/2024
Initial Date FDA Received02/08/2024
Supplement Dates Manufacturer Received03/12/2024
Supplement Dates FDA Received04/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-