Model Number RT226 |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/12/2024 |
Event Type
malfunction
|
Event Description
|
A distributor reported on behalf of a healthcare facility, via a fisher & paykel (f&p) healthcare representative in japan that the water trap of a rt226 infant ventilator breathing circuit was observed to be cracked when being removed from the packacking.There was no patient involvement reported.
|
|
Manufacturer Narrative
|
(b)(4).The subject rt226 infant ventilator breathing circuit has been requested to be returned to f&p healthcare new zealand for evaluation.We will provide a follow up report upon completion of our investigation.
|
|
Manufacturer Narrative
|
(b)(4).Method: the subject rt226 infant ventilator breathing circuit was returned to f&p healthcare in new zealand where it was visually inspected and analysed.Results: visual inspection of the returned rt226 infant ventilator breathing circuit confirmed a long vertical crack on the outside of the water trap.The inside wall of the water trap was observed cracked in multiple directions.Material analysis confirmed that there was no evidence of material degradation and the most likely cause of the crack is an external mechanical impact.Conclusion: we are unable to determine the cause of the reported crack.However, based on the conducted analysis, the crack is likely to have occurred due to an external impact.All rt226 infant low flow breathing circuits are visually inspected, and pressure and flow tested during production, and those that fail are rejected.The subject breathing circuit would have met the required specifications at the time of production.The user instructions that accompany the rt226 infant low flow breathing circuit state the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
|
|
Event Description
|
A distributor reported on behalf of a healthcare facility, via a fisher & paykel (f&p) healthcare representative in japan that the water trap of a rt226 infant ventilator breathing circuit was observed to be cracked when being removed from the packaging.There was no patient involvement reported.
|
|
Search Alerts/Recalls
|