C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 1806060J |
Device Problems
Device Damaged Prior to Use (2284); Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/15/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4: (expiry date: 07/2024).H11: section a: through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that prior to a port placement, there was allegedly a seal tear inside in the actual item before opening the package.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one sealed powerport isp mri implantable port kit was returned for sample evaluation.Visual evaluation was performed.The outer plastic packaging sealed with no damages was noted.Tyvek label noted to be lifted and torn in the upper right-hand corner.As per manufacturing review confirmed the seal tear and upper left corner was slightly raised and tyvek label of the product was missing a part of the seal (adhesion).No other anomalies were observed.Therefore, the investigation is confirmed for reported seal tear prior to use issues.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiry date: 07/2024), h6 (device).H11: h6 (method, result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that prior to a port placement procedure, there was allegedly a seal tear in the actual item before opening the package.There was no patient contact.
|
|