Model Number N/A |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/11/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during a knee procedure, while the wound was being irrigated, the patella packaging was opened.Moisture was noted in the sterile packaging but it is unknown whether it was already in the packaging or due to the irrigation.Another patella was used to complete the procedure without issue.
|
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event could not be confirmed.Visual evaluation of the returned product found no moisture or staining in or on the packaging.It has been reported that the moisture may have been introduced during the irrigation step during the surgery.Device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for the reported part and part lot combination.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|