Catalog Number B12LT
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 2/9/2024.D4: batch # unk.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during an unknown procedure, duckbill valve fell off.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Manufacturer Narrative
(b)(4).Date sent: 3/5/2024 d4: batch # a9dz1u investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the b12lt device was received with the duckbill out of position and present.In addition, the duckbill was returned inside a plastic bag.However, it is known from the history of the device that the condition of the duckbill out of position may lead to insufflation issue.No conclusion could be reached as to what might have caused the duckbill out of position.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18676135
MDR Text Key335949782
Report Number3005075853-2024-01185
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB12LT
Device Lot NumberBATCH:A9DZ1U
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/06/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received02/09/2024
Supplement Dates Manufacturer Received02/14/2024
Supplement Dates FDA Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-