The hospital reported that ultima activator ii reusable drive mech had corrosion on surface.The sterilization process was the following: first, mechanical cleaning and disinfection.Second chemical use of (b)(6) mediclean advanced.Third, thermal disinfection a0 value 3000.Fourth, steam sterilization 134° c - 5 minutes.There was no patient involvement.
|
Trackwise#: (b)(4).Updated sections: b4, d9, g3, g6, h2, h3, h6, h10, h11.Corrected sections: d4--serial # corrected from "(b)(6)" to "asku" as the serial number is unknown.The device was returned for evaluation on 02/29/2024.Photographs were provided by the account.A photographic evaluation was conducted.Signs of clinical use and no evidence of blood was observed.Corrosions was observed on the device.There were no other visual defects observed.An investigation was conducted on 03/05/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood was observed.Corrosions was observed on the device.Based on the photographic evaluation, as well as the evaluation results., the reported failure "corroded" was confirmed.The reported device is an oem device, therefore, a lot history review was not applicable.A lot/serial number was not provided and the specific product lot/serial number cannot be identified from a ship history, therefore it is impossible to obtain a certificate of conformance.
|