It was reported that catheter perforation occurred.The target lesion was in the lower extremity vein.An angiojet solent omni catheter was selected for lower extremity venous thrombectomy procedure.After the device was primed and flushing was performed, it was noted that the guidewire pierced from the side of the catheter and the body of the device was broken, affecting normal use.The procedure was completed with another of the same device.No complications reported and the patient was stable.
|
Device evaluated by mfr: the device was returned for evaluation.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The catheter shaft showed multiple bends and kinks.During the disinfecting process by the lab assistants, it was noted the outer shaft was barely attached and unintentionally separated while wiping the device.Functional testing was not attempted due to the extreme damage to the device.During analysis, it was observed the shaft and hypotube were completely separated at 114 cm from the tip.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The complaint was confirmed for under issues related to a hypotube and shaft separation.Shaft kinks were also confirmed.E1.Initial reporter facility name-(b)(6).
|
It was reported that catheter perforation occurred.The target lesion was in the lower extremity vein.An angiojet solent omni catheter was selected for lower extremity venous thrombectomy procedure.After the device was primed and flushing was performed, it was noted that the guidewire pierced from the side of the catheter and the body of the device was broken, affecting normal use.The procedure was completed with another of the same device.No complications reported and the patient was stable.
|