This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The device was returned to olympus for inspection, and the customer's complaint was confirmed.In addition, the following non-reportable malfunctions were found during the device evaluation: front panel is damage, customer installed wrong lamp and lamp meter reads over 500 hours.Based on the results of the investigation, a definitive root cause could not be determined.However, it is like that the event reported as "image was intermittently interrupted" was caused by a failure by the effect of installing the wrong lamp or not being able to connect properly due to wear on the scope connector socket.Olympus will continue to monitor field performance for this device.
|