• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500; BLOOD GAS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500; BLOOD GAS MONITOR Back to Search Results
Model Number CDI510H
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2024
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.H3: 81 - evaluation is in progress, but not yet concluded.
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime a leakage was observed.As per the subsidiary, during priming, leaks were observed at both ports of the shunt sensor.The shunt sensor was not use and a backup shunt sensor was used without problems.No patient involvement, product was changed out, procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on february 09, 2024.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 11, 3331, 4210, 19).The returned sample was visually inspected upon receipt, during which it was noted that the sparger assembly and white luer cap were not present.A white luer cap was added to the sample.It was then leak tested, as received by connecting with a calibrated manometer, submerged into a water bath, and pressurized up to 1030 mmhg and a leak was noted immediately.The large bore adapter blue cap was then loosened and re-tightened by hand.The sample was then leak tested a second time by connecting with the calibrated manometer, submerged into a water bath, and pressurized up to 1030 mmhg, and no leaks were noted.A retention sample from the same lot number was visually inspected and confirmed to have no traces of buffer on the outside of the unit or inside the pouch.It was then leak tested the way, and no leaks were noted on the retention sample.The root cause for this event was determined to be the large blue vent cap for shunt sensor was not fully tightened either during setup of the circuit, or after the gas calibration.When the large blue vent cap was loosened, it had not been re-tightened fully, causing a leak from the cap.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNT SENSOR SYS500
Type of Device
BLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
125 blue ball road
elkton, MD 21921
7346634145
MDR Report Key18681223
MDR Text Key335109740
Report Number1124841-2024-00023
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDI510H
Device Catalogue NumberN/A
Device Lot Number3M09G
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/18/2024
Initial Date FDA Received02/09/2024
Supplement Dates Manufacturer Received02/19/2024
Supplement Dates FDA Received03/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-