• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW ARTERIAL CATH. SET: 22 GA X 5CM; CATHETER PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW ARTERIAL CATH. SET: 22 GA X 5CM; CATHETER PERCUTANEOUS Back to Search Results
Catalog Number SAC-00522-PBX
Device Problem Unraveled Material (1664)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported "the swg tangles on the needle, and when trying to move backwards the swg it broke".No patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
 
Event Description
It was reported "the swg tangles on the needle, and when trying to move backwards the swg it broke." no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine.".
 
Manufacturer Narrative
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.The ifu was reviewed, and it informs the user not to retract the guide wire against the needle bevel.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW ARTERIAL CATH. SET: 22 GA X 5CM
Type of Device
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
mariah mackinnon
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18681375
MDR Text Key335084145
Report Number3006425876-2024-00106
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K171146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberSAC-00522-PBX
Device Lot Number71F22L1361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/22/2024
Initial Date FDA Received02/09/2024
Supplement Dates Manufacturer Received03/12/2024
Supplement Dates FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NONE REPORTED; NONE REPORTED
-
-