Model Number DSX500T11C
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/02/2024
Event Type  malfunction  
Event Description
A dreamstation auto cpap device was returned to a third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and blower foam degradation was found.During the evaluation, foam particles were observed inside the blower kit.In addition, there was corrosion in the connector.The device was scrapped.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18683048
MDR Text Key335073563
Report Number2518422-2024-06803
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberDSX500T11C
Device Catalogue NumberDSX500T11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2022
Initial Date Manufacturer Received 01/02/2024
Initial Date FDA Received02/12/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/04/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1974-2021
Patient Sequence Number1
-
-