MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW ARTERIAL CATH SET: 20 GA X 5CM; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW ARTERIAL CATH SET: 20 GA X 5CM; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number SAC-00520
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/15/2024
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).The customer provided one photo for analysis.The complaint of a guide wire kinked in package was confirmed through visual inspection of the provided photo.The customer also returned one unopened sac kit for evaluation.No signs of use were observed.The kit was opened to further inspect the returned components.Visual inspection of the guide wire revealed one kink in the guide wire body.Microscopic examination confirmed the damage.Both welds are present and appeared full and spherical.Creasing and bending was also observed on the guide wire tube and the product packaging.The lidstock clearly states "do not bend".The kink in the guide wire measured 73mm from the distal end.The guide wire length measured 350mm which is within the specification of 345-355mm per product drawing.The guide wire outer diameter measured 0.52mm which is within the specifications of 0.508-0.533mm per product drawing.Functional inspection of the guide wire was performed per the instructions for use (ifu) provided with the kit which states, "insert desired tip of spring-wire guide through introducer needle into artery (until depth-marking [if provided] on wire enters hub of needle).Thread tip of indwelling catheter over spring-wire guide." the undamaged portions of the guide wire were threaded through the returned 20 ga introducer needle with no resistance.A manual tug test confirmed the distal and proximal welds were attached.The manufacturing site was previously consulted regarding the observed failure mode for related complaints.They indicated that the observed kink, in addition to the creasing in the packaging, is consistent with damage caused by shipping and handling.As part of the guide wire manufacturing process, each guidewire is passed through a ring gauge so it is unlikely that this defect would have been present prior to release from the manufacturing site.A device history record review was performed with no relevant findings.The ifu provided with this kit warns the user , "precaution: use of excessive force may damage catheter or guidewire.Do not use if package is damaged." and engineering change order (eco) was implemented in april 2017 to update the product labeling to clearly inform the customer to not bend the product and is intended to reduce the potential for product damage.This product was manufactured (february 2022) after the implementation date of the eco.The lidstock clearly states, "do not bend." the customer report of a kinked guide wire prior to use was confirmed through complaint investigation of the returned sample.One kink was observed on the guide wire body.In addition, several folds and creases were observed on the outside packaging.The guide wire met all relevant dimensional requirements, and a device history record review was performed with no relevant findings.The lidstock clearly states "do not bend." based on the customer description and the sample received, storage and shipping caused or contributed to this event.Teleflex will continue to monitor and trend for complaints of this nature.
 
Event Description
It was reported that "we found the product with damaged packaging inside the box during receipt and labelling inspection." the issue was identified at incoming inspection /inventory review.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW ARTERIAL CATH SET: 20 GA X 5CM
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
elaine cully
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18683485
MDR Text Key335073230
Report Number3006425876-2024-00196
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K093050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSAC-00520
Device Lot Number71F22B1545
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2024
Initial Date Manufacturer Received 01/15/2024
Initial Date FDA Received02/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED
-
-