• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORPORATION FIRSTPASS MINI STRAIGHT; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORPORATION FIRSTPASS MINI STRAIGHT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number 72290128
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Manufacturer Narrative
H10: internal complaint reference: case(b)(4).
 
Event Description
It was reported that during arcr procedure, the first pass mini's capture window of upper jaw was broken.The broken pieces were removed from the patient.The procedure was completed with a smith and nephew back up device and it is unknown if there was a surgical delay.No further complications were reported.
 
Manufacturer Narrative
The reported device was received for evaluation.A visual evaluation showed the device was not returned in any original packaging.The suture capture has been disconnected from the upper jaw.The returned suture capture is not deformed or broken.No other visual deficiencies.A functional evaluation showed pulling the lever will close the jaw and deploy the suture passer needle.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that could have contributed to the failure include rough handling or excessive force to the device no containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIRSTPASS MINI STRAIGHT
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18684917
MDR Text Key335187438
Report Number3006524618-2024-00062
Device Sequence Number1
Product Code NBH
UDI-Device Identifier00885556694558
UDI-Public00885556694558
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72290128
Device Lot Number2121090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/23/2024
Initial Date FDA Received02/12/2024
Supplement Dates Manufacturer Received03/12/2024
Supplement Dates FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-