MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9735669 |
Device Problem
Device Sensing Problem (2917)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Section d references the main component of the system.Other medical products in use during the event include: product id h3, h6: a medtronic representative went to the site to test the equipment.Testing revealed that the axiem flat plate emitter was not picking up the trackers across the board.The fov seemed to be smaller than intended.Codes b01, c07 and d02 are applicable to this analysis. medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess) procedure.It was reported that the site was unable to navigate the suction.It verified fine but then when bringing it into the field it would flicker red and they would lose tracking when placing the instrument into the patient. the site was unable to get the system to perform as intended and elected to continue without navigation.Technical services (ts) also had the site check extensively for metal interference.After changing the instrument tracker they were still experiencing issues tracking the straight suction.The instrument did verify, but the tracking was intermittent when attempting to navigate in the sinus. ts had the site attempt to adjust the flat emitter and the issue persisted. the site also began to have the same issue with the ostium seeker. the site attempted to replace the gel pad on top of the emitter with a "donut".The issue persisted. upon checking the tracking details, the non-invasive patient tracker (nipt) had a geometry error of 1.15 and the instruments were all between 3 and 4.The site replaced the patient tracker, was able to register successfully but continued to have issues tracking other instruments. the site noted that the nipt was about 7-8 inches above the flat emitter and that the instruments only seemed to track when the instrument tracker was level with the patient's face.This resulted in a surgical delay of less than one hour.There was no impact on patient outcome.
|
|
Manufacturer Narrative
|
Additional information in section a.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id 9733752 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3, h6: product 9733752, lot number: 603006334 was received by the manufacturer for analysis.Reported problem couldn't be duplicated.Flat emitter was connected to the "lat" station and passed a burn-in test over the weekend (72hr) without issues.C19 and d14 are applicable.Previously reported code b01 is also applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|