C.R. BARD, INC. (BASD) -3006260740 ACCUCATH ACE INTRAVASCULAR CATHETER 22GX1.25; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Break (1069)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 01/18/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H11: section a: through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant, reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.
|
|
Event Description
|
It was reported by customer that when attempting to retract guidewire resistance was felt.Rn suspected guidewire was rubbing against the inside of the catheter attempted to withdraw catheter with guidewire.There was a difference in length when comparing the used catheter to a new unused one.Ultimately, the tip, coil at the top was not attached to the used catheter which resulted in patient being sent to the operating room to have it surgically removed.
|
|