Model Number P6881 |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/02/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The baxter technician found the blower power cord needed to be replaced.Per the hillrom service manual inspect power cord for damage, twisting and replace if necessary.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in (b)(6) 2023.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the blower power cord to resolve the reported event.Based on this information, no further action is required.
|
|
Event Description
|
Baxter received a report from a baxter technician stating the bed blower power cord was damage with exposed copper wires.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
|
|