ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0702-XTW
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Event Description
It was reported this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 4+, flail leaflet, and a pre-existing chordal rupture.During preparation of an xtw clip (30627r1030), it was observed that one of the grippers was not retracting.The device was replace and an xtw clip was successfully implanted.To further reduce mr, an additional xtw clip (30912r1068) was implanted, imaging was poor.After two minutes, the second implanted clip had detached from the anterior leaflet and remained attached to the posterior leaflet (single leaflet device attachment/slda).Leaflet damage was observed after the slda.Mr was reduced to a grade of 3+, but for treatment, the physician decided to perform mitral valve replacement.
 
Manufacturer Narrative
All available information was investigated, and the reported difficult to open or close (gripper actuation - single) could not be confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, a cause of the reported difficult to open or close (gripper actuation - single) could not be determined as the issue was not identified in device analysis.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18695246
MDR Text Key335436331
Report Number2135147-2024-00626
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0702-XTW
Device Lot Number30627R1030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2024
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-