MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number B35200 |
Device Problems
Break (1069); High impedance (1291); Material Deformation (2976)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 07/10/2023 |
Event Type
Injury
|
Event Description
|
It was reported that the manufacturer representative (rep) received an email today that has been looping around with the neurologist dated 7/10/2023.Caller reported patient was having impedance issue with electrodes: left hemisphere: 1a, 2b, 3.Right hemisphere: 9b.Caller reported those contacts are showing >10k ohms.Caller reported xray was performed, did not show any hardware issue.Caller reported patient feeling fine, no changes in her therapy.Caller reported neurologist did not want anything change as patient was doing well.Impedance were re-tested the week of (b)(6), 2024, impedance went from 10k ohms down to 5k ohms, same electrodes: 1a, 2b, 3, and 9b.Caller reports patient felt her therapy has diminished, no shocking sensation.Caller reports patient is scheduled to be seen by the neurologist on monday.Physician will be palpating the area with impedance testing to see if there is any changes with impedance and palpation.
|
|
Manufacturer Narrative
|
Section d references the main component of the system.Other medical products in use during the event include: medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3.Analysis of the extension (s/n (b)(6)) found the conductor on the proximal end of the conductor was broken.Analysis of the extension (s/n va2mn6ev19) found the conductor on the proximal end of the conductor was broken.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the manufacturer¿s representative (rep), which was confirmed with the healthcare provider (hcp), reported the cause of the impedance issue wasn¿t determined.Values ranged from >5,000 on monopolar to >10,000 on the left and right as well as high.A shunt x-ray was performed which showed the system seemed to be in good condition, however it was limited views.Surgery was scheduled to explore integrity of the lead, extension, and neurostimulator.
|
|
Manufacturer Narrative
|
Section d10 references: product id b3400060m , serial# (b)(6), product type extension.Product id b3400060 , serial# (b)(6), product type extension.Product id b32000, lot# 082m09323 , implanted: (b)(6) 2023, product type accessory.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the manufacturer¿s representative (rep) reported the physician performed a revision on (b)(6).During that time they determined the right extension with the marker¿s outer insulation was compromised that could be pulled apart to expose the lead coils inside.There was nothing apparent the left extension although it seemed to have kinks.In addition, they experienced a system error with a lost connection twice today (b)(6) so they had to put the communicator into the field to allow for a connection.
|
|
Manufacturer Narrative
|
H3.Analysis of the extension (s/n (b)(6)) found the extension/body/outer insulation was separated at the tube overlap joint.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|