Model Number BREV100
Device Problem Failure to Obtain Sample (2533)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2024
Event Type  Injury  
Event Description
It was reported that during a brevera procedure on (b)(6) 2024, the biopsy was not able to be completed as the needle was inside the breast.Case had to be cancelled.A field engineer examined the equipment and the reported problem could not be duplicated.Driver was cycled 12 times with no issues , logs gathered and submitted to support for review.No definitive cause could be established and the system was operating per manufacturer´s specifications.
 
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BREVERA BREAST BIOPSY SYSTEM
Type of Device
INSTRUMENT, BIOPSY
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free zone building b24
san jose 20102
CR   20102
MDR Report Key18697852
MDR Text Key335293078
Report Number1222780-2024-00058
Device Sequence Number1
Product Code KNW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberBREV100
Device Catalogue NumberBREV100
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-