• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Flaked (1246)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/30/2024
Event Type  malfunction  
Event Description
A facility reported that after they used the perforator id (b)(6) they noticed that the perforator label had been rubbing against the portion of the scalp that had been pulled back to allow them to prepare for the craniotomy.The tissue had blood on it and the bone was being irrigated during the procedure, so the combination of blood and irrigation eventually moistened the sticker and the friction caused by rubbing against the exposed tissue led to part of the sticker rubbing away.This did not affect the procedure.The perforator was stored on a cart in the hallway of the operating room, which is where they have stored the perforators for a while.It is unknown if any label parts fell into the surgical site.Due to the friction, it is possible that very small pieces might have, but they may have been removed with irrigation during and after the procedure.
 
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18699014
MDR Text Key336407347
Report Number3014334038-2024-00030
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number261221
Device Lot Number7257266
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/30/2024
Initial Date FDA Received02/13/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age41 YR
Patient SexFemale
Patient RaceWhite
-
-