E2, e3- information has been requested but unknown at this time.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the suggested event was confirmed; however, a root cause could not be identified.The probable cause was likely due to use error.The instruction manual of cyf-v2 identifies the following related verbiage about reprocessing method which could have prevented the phenomenon: ¿ chapter 6 compatible reprocessing methods and chemical agents ¿ chapter 7 cleaning, disinfection, and sterilization procedures olympus will continue to monitor field performance for this device.
|