• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE C; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE C; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problems Loosening of Implant Not Related to Bone-Ingrowth (4002); Migration (4003)
Patient Problems Ossification (1428); Pain (1994); Loss of Range of Motion (2032); Ambulation Difficulties (2544)
Event Date 01/17/2024
Event Type  Injury  
Manufacturer Narrative
Cmp-(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event Description
It was reported that the patient underwent a knee revision approximately 5 years post implantation due to pain, instability, decreased rom, difficulty ambulating, and tibia subsidence.During the revision, heterotopic ossification, confirmed subsidence, and loosening of tibia was noted.
 
Manufacturer Narrative
No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: office visit: ongoing pain with activities and instability, pain 8 on 1-10 scale, difficulty with stairs, dealing with chairs, shoes, and prolonged walking with antalgic gait.Rom: -5-80 degrees.Revision report: tibia found to be subsided and loose.Explanted with ease including the bulk of cement.A definitive root cause cannot be determined.Complaint is confirmed with the provided medical records.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE C
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18699950
MDR Text Key335293946
Report Number3007963827-2024-00037
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172524
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42532006402
Device Lot Number63987510
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/13/2024
Supplement Dates Manufacturer Received04/17/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/12/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight64 KG
-
-