Model Number 560BC |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/15/2024 |
Event Type
malfunction
|
Event Description
|
Medtronic received information that prior to use of a bio-console instrument, it was reported that the instrument was indicating an erratic blinking light for charging.The customer performed a battery replacement and the instrument still had an erratic blinking light for the battery indication.The instrument was replaced.There was no patient involvement, so no adverse effect occurred.
|
|
Manufacturer Narrative
|
Device evaluation summary: the reported issue that the instrument was indicating an erratic blinking light for charging despite a battery replacement was verified during service.The service technician tested the new batteries with no change.The service technician performed a system calibration of the power supply and the main pcb with no change.The issue was resolved by replacing the power supply, calibrating the instrument and verifying that the instrument started charging properly through ad50, ad60, and ups100 tests.Post-repair testing was performed per specifications.Note: the instrument was analyzed in the facility by a medtronic field service technician.The instrument did not return to a medtronic facility for analysis/service.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|