ZIMMER BIOMET, INC. RINGLOC BI-POLAR 28X51MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).D10: cat# 00877502803 lot# 3180964 biolox⮠delta, ceramic femoral head, l, 㸠28/+3.5, taper 12/14.The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the metal cup could not be placed correctly over the liner.After the liner was placed on the head and snapped into place, the cup was put onto the liner and became stuck at an angle.Force was used and the pieces would not go together.A second set of implants was used to complete the procedure.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.One ringloc bi-polar 28x51mm item# 11-165226 lot# 65991003 was returned and evaluated.Upon visual inspection the head and liner have been assembled and could not be removed from the shell.The assembly is sitting at an angle inside of the shell and could not be removed.No measurements were taken due to the device being assembled and not being able to be separated.A review of the device history records identified no related deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.This complaint was confirmed based on the returned devices, as the assembly was sitting at an angle and could not be removed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RINGLOC BI-POLAR 28X51MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18703353
MDR Text Key335361662
Report Number0001825034-2024-00390
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051569
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number11-165226
Device Lot Number65991003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/29/2024
Initial Date FDA Received02/14/2024
Supplement Dates Manufacturer Received06/10/2024
Supplement Dates FDA Received06/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE
Patient SexMale
-
-