• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/11/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned and the evaluation and the customer's allegation was confirmed.The evaluation found damage to the connector of the manifold unit.Based on the results of the investigation, it is likely the damaged connector led to the malfunction.However, a definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.
 
Event Description
It was reported, the high flow insufflation unit outflow is not measured correctly.The issue occurred during installation, while using the device in a therapeutic cholecystectomy procedure.The procedure was completed using the subject device.There was no patient harm or impact associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18703470
MDR Text Key336525620
Report Number3002808148-2024-01374
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/11/2023
Initial Date FDA Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-