MEDTRONIC XOMED INC. IPC® HANDPIECE - INDIGO¿ DRILL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
|
Back to Search Results |
|
Model Number 1845000 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H3: analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
A health care professional reported that, indigo handpiece motor was getting warm.There was no patient impact.Additional information received that during intra-op surgery has been completed with another indigo device with drilling rmm limits, and it was used in forward drilling mode.Different irrigation was used depending on the stage in the surgery.
|
|
Manufacturer Narrative
|
H3: product analysis found that, the customer's complaint has not been confirmed.The motor was not overheating, but the collet does.Pre-repair temperature in fahrenheit in 1 min of testing handpiece temperature were t1-80 and t2-86.H6: codes are updated.Previously applied fdm b21, fdr c21, fdc d16 and img g04063 codes no longer applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|