Model Number NEU_EPIDURALNEEDLE |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/06/2023 |
Event Type
malfunction
|
Event Description
|
Information was received from an unknown source regarding a patient who was implanted with an implantable neurostimulator (ins).It was reported stylet wouldn¿t remove smoothly for needle.The doctor tested this before it touched the patient.The needle was not used on patient or any part of the procedure.Rep grabbed a new kit and gave him a new needle.
|
|
Manufacturer Narrative
|
Product analysis #(b)(4): analysis information -- (b)(6) 2024 11:13:19 cst pli# 10 product id# neu_epiduralneedle below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and functional testing.Analysis determined that there was a sharp edge on the undercut side of the spoon of the needle.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
A correction was made to update analysis results of pli 10.H3: pli# 10 product id# neu_epiduralneedle found no anomaly.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
2023-oct-06 mpxr 1110464 (): information was received from an unknown source regarding a patient who was implanted with an implantable neurostimulator (ins).It was reported stylet wouldn¿t remove smoothly for needle.The doctor tested this before it touched the patient.The needle was not used on patient or any part of the procedure.Rep grabbed a new kit and gave him a new needle.2023-11-01 e1, e2, e3, rp (rep): no new information.
|
|
Search Alerts/Recalls
|
|
|