MAUDE Adverse Event Report: FARAPULSE, INC. FARAWAVE PULSED FIELD ABLATION CATHETER, 31MM; CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FARAPULSE, INC. FARAWAVE PULSED FIELD ABLATION CATHETER, 31MM; CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER Back to Search Results
Lot Number 0008021707
Device Problem Entrapment of Device (1212)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 02/06/2024
Event Type  Injury  
Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported that during an irreversible electroporation ablation procedure to treat atrial fibrillation (a fib) using a farawave catheter the guidewire became stuck in the guidewire lumen.After attempting ablation on the right inferior pulmonary vein (ripv) they attempted to retract the guidewire but found it to be stuck.They removed the catheter and guidewire through the sheath.They were able to remove the guidewire from the catheter and observed tissue in the tip of the guidewire lumen.They exchanged the catheter, and the issue was resolved.The procedure was then completed with no patient complications.The catheter is expected to be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FARAWAVE PULSED FIELD ABLATION CATHETER, 31MM
Type of Device
CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER
Manufacturer (Section D)
FARAPULSE, INC.
3715 haven avenue
suite 110
menlo park CA 94025
Manufacturer (Section G)
AGILE MV INC
3700 st patrick st suite 102
montreal, qc H4E 1 A2
CA   H4E 1A2
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key18707490
MDR Text Key335405281
Report Number2124215-2024-08783
Device Sequence Number1
Product Code QZI
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Lot Number0008021707
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/14/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-