MAUDE Adverse Event Report: BIOMET UK LTD. DELTA CERAMIC FEM HD 36/-3MM; HIP PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. DELTA CERAMIC FEM HD 36/-3MM; HIP PROSTHESIS Back to Search Results
Catalog Number 650-0660
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/22/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that during a procedure, the surgeon noticed scuff/scratch marks on the newly opened ceramic femoral head.The surgery was completed with a substitute ceramic femoral head.This event is related to a malfunction that could potentially lead to a sterility issue, serious injury.However, no patient harm or further outcome was reported.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, d9, g3, g6, h2, h3, h6, h10.Visual examination of the returned product identified that the item and lot numbers match what is on the complaint.Inner tyvek has been tampered with and so was the box.Scuffs and scratches noticed on the part.Inside the taper noticed scratches and a stain in the bottom of the taper.Shrink wrap and box have already been opened.A review of the device manufacturing records confirmed no abnormalities or deviations.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA CERAMIC FEM HD 36/-3MM
Type of Device
HIP PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18710157
MDR Text Key335743940
Report Number3002806535-2024-00055
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200959
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number650-0660
Device Lot Number3168015
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/22/2024
Initial Date FDA Received02/15/2024
Supplement Dates Manufacturer Received03/15/2024
Supplement Dates FDA Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-