• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON Back to Search Results
Model Number 97800
Device Problems High impedance (1291); Unstable (1667); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2024
Event Type  Injury  
Manufacturer Narrative
Other medical products in use during the event include: brand name: surescan, product id: 978b128 (lot: va2l7fn), product type: 0200-lead, implant date: (b)(6) 2022, explant date: (b)(6) 2024.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information was received from a manufacturer representative (rep) regarding a patient who was implanted with an implantable neurostimulator (ins) for fecal incontinence and urinary/bowel dysfunction.It was reported that when the rep saw the patient they discovered that all impedances were greater than 4000 ohms.It was noted that the patient had not had any falls or car accidents.A lead revision was scheduled for (b)(6).
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a manufacturer representative (rep).The rep reported that the cause of the high impedance was determined.During the surgery on (b)(6) 2024 the device appeared to have flipped many times in the pocket, causing the lead to coil up again and again.The lead and ins were removed and a new ins and lead were implanted.They confirmed the issue was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM X
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18711094
MDR Text Key335469474
Report Number3004209178-2024-04806
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00763000484668
UDI-Public00763000484668
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2023
Device Model Number97800
Device Catalogue Number97800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/08/2024
Initial Date FDA Received02/15/2024
Supplement Dates Manufacturer Received02/16/2024
Supplement Dates FDA Received03/08/2024
Date Device Manufactured05/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age82 YR
Patient SexFemale
-
-