SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/30/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned and the evaluation found that the device could not be lit occasionally due to defective power unit and igniter.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the light source main lamp was not on.The issue occurred during reprocessing.The patient was not under sedation.There was no delay, and there were no reports of patient harm.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation, as well as corrections to e1, e2 and e3.The information in e1, e2 and e3 was inadvertently omitted from the initial medwatch report.   new information added in fields: h4 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.   based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the device could not be lit occasionally, due to a power unit failure and ignitor failure.Olympus will continue to monitor field performance for this device. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18712143
MDR Text Key336410624
Report Number3002808148-2024-01423
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/30/2024
Initial Date FDA Received02/15/2024
Supplement Dates Manufacturer Received03/01/2024
Supplement Dates FDA Received03/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/18/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-