|
Catalog Number 72290128 |
Device Problems
Unstable (1667); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/24/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during an arthroscopy, when the tissue was punctured with the firstpass, the catch mechanism came loose, nothing remained in the patient.The procedure was completed with a s+n back up device.There was a delay less than 30 minutes and no further complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference case (b)(4).
|
|
Manufacturer Narrative
|
H6 codes (medical device problem code) updated.H3, h6: the reported device was received for evaluation.A visual evaluation showed the device was not returned with any original packaging.The suture capture has been disconnected from the upper jaw and is deformed.Bio debris is present.No other visual deficiencies.A functional evaluation showed pulling the lever will close the jaw and deploy the suture passer needle.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.An analysis of the customer provided images found that the suture capture looks displaced from its normal position.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include (1) excessive force, (2) tissue thickness, (3) damage or debris on the device tip between passes.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|