ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; Pulse generator, permanent, implantable Back to Search Results
Model Number PM2240
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 02/14/2024
Event Type  malfunction  
Event Description
This report is to advise of an event observed during analysis.
 
Manufacturer Narrative
The device was returned without identified device or use problem.A malfunction based on analysis was found.As received, the device output and telemetry communication were normal.Visual inspection of the header attachment area detected a bonding anomaly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The device was tested for a hermeticity breach which was not observed.The device was cut open to enable further testing and battery was found in normal range.The hybrid circuitry was tested, and the results indicated normal current drain.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the assurity and endurity pacemakers header anomaly advisory issued by abbott on 15 march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18712908
MDR Text Key335840854
Report Number2017865-2024-32766
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2017
Device Model NumberPM2240
Device Lot NumberA000024319
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/14/2024
Initial Date FDA Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1530-2021
Patient Sequence Number1
-
-