Catalog Number UNK_WWA |
Device Problems
Break (1069); Fracture (1260); Osseointegration Problem (3003); Insufficient Information (3190)
|
Patient Problems
Cyst(s) (1800); Failure of Implant (1924); Inadequate Osseointegration (2646); Insufficient Information (4580)
|
Event Date 01/22/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
The device is not available for evaluation as it remains implanted in the patient.A review of the device history is not possible because the lot number was not communicated.If additional information becomes available, it will be provided on a supplemental report.H3 other text : device remains implanted in patient.
|
|
Event Description
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery for reasons that are not available at the time of this report.
|
|
Manufacturer Narrative
|
Correction - h6 (results code and conclusion code).The reported event could be confirmed since images of ct scans shows the talar component is loosened and one of the pegs is broken.Upon further investigation of the ct scans by healthcare professionals the following was observed: ¿there is some radiolucence and some cysts adjacent to the talar component.Loosening is likely, there is no clear sign of migration.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the catalog number and lot number were not communicated.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
|
Event Description
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery for reasons that are not available at the time of this report.
|
|