Model Number 5584111 |
Device Problems
Break (1069); Use of Device Problem (1670)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/22/2024 |
Event Type
malfunction
|
Event Description
|
Information was received from a healthcare professional via a manufacturer representative regarding a patient with pre-operative di agnosis for l4-5 posterior fusion.Procedure or technique used implanting pedicle screws at l5.Levels implanted l4-5.It was reported that the driver was being used to insert a 10.5mm screw.The surgeon had under-tapped using an 8.5mm tap as he was worried about having a good tight fit.Bone was very sclerotic and hard, and the driver was difficult to turn.Had to use significant force to turn it, and the tip broke off in the head of the screw.There was no fragment left inside the patient.There was no patient symptoms or complications as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3.Product analysis product id # 5584111; lot no.# sw23c018 visual inspection revealed the tip of the driver has been broken.Damage present is consistent with torsional overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|