It was reported that a patient presented with grade 4 functional mitral regurgitation (mr) and a prolapsed anterior leaflet for a mitraclip procedure.All devices were prepped and executed according to instructions for use (ifu).While pushing the clip introducer of the clip delivery system (cds) through the hemostasis valve of the steerable guide catheter (sgc), it was noted that the fluid level in the hemostasis valve was diminishing.The clip with the clip introducer were removed from the hemostasis valve and the column held.The clip introducer was inspected and a leak was found at the blue opaque tip, which caused the loss of fluid column at the hemostasis valve.The clip was replaced and a new clip was implanted without any problems.The patient is doing well.The mr was reduced to grade 1.There were no adverse patient effects or clinically significant delay.
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation was unable to determine a cause for the reported leak.There is no indication of a product issue with respect to manufacture, design, or labeling.
|