MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735665
Device Problems Application Program Problem (2880); Unintended Application Program Shut Down (4032)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2024
Event Type  malfunction  
Manufacturer Narrative
H6) no parts have been received by the manufacturer for evaluation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess) procedure.It was reported that error 64 was displayed while the registration 3d model is being built.It then returned them to the log in screen.This resulted in a surgical delay of less than one hour.There was no impact on patient outcome.
 
Manufacturer Narrative
D10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735736, version 2.0.0 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3: logs were received and software analysis was completed.Logs captured couple of corefiles and segfaults in "trualgorithmservice" service during fess procedure while registration task on the date of the issue reported.The core was generated by `trualgorithmservice' and the program terminated with signal sigsegv, segmentation fault during "malloc ()" call which was triggered by "tru::highcu rvpointsclusterslocals::nighbpoints()" function.The syslog's also captured segfaults in "trualgorithmservice" at library "libc-2.19.So".A software issue was confirmed.The reported event results from a software malfunction.Codes b01, c10, d18 are applicable to this analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION¿ S8 SYSTEM
Type of Device
EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18721312
MDR Text Key336179872
Report Number1723170-2024-00572
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/16/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/06/2024
Initial Date FDA Received02/16/2024
Supplement Dates Manufacturer Received03/06/2024
04/25/2024
Supplement Dates FDA Received03/06/2024
05/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/04/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SEE H11
-
-