Model Number 7122Q/65
Device Problems Failure to Capture (1081); High impedance (1291); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2024
Event Type  malfunction  
Event Description
It was reported the patient presented for implant procedure.During surgery, it was discovered the right ventricular lead exhibited a loss of capture, a loss of sensing, and high pacing impedance.The physician elected to complete the procedure with a different lead.The patient was in stable condition.
 
Manufacturer Narrative
The reported events of failure to sense, high pacing impedance and failure to capture were not confirmed.A complete lead was returned in one piece.Visual and x-ray examination of the lead did not find any anomalies.Electrical testing did not find any indication of conductor fractures or internal shorts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18721581
MDR Text Key336000458
Report Number2017865-2024-32796
Device Sequence Number1
Product Code NVY
Combination Product (y/n)Y
Reporter Country CodeCO
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/65
Device Lot NumberS000091586
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/03/2024
Initial Date FDA Received02/16/2024
Supplement Dates Manufacturer Received03/25/2024
Supplement Dates FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-