Model Number 45031
Device Problems Break (1069); Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Event Description
It was reported that tip of the catheter was hydroped and ruptured occurred.The target lesion was located in the lower extremity vein.An angiojet solent omni was selected for use for thrombectomy procedure.During the procedure, error was reported and the tip of the balloon was hydroped and ruptured.The procedure was successfully completed with another of the same device and the patient was stable.
 
Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an angiojet solent omni catheter.Inspection of the device revealed shaft kinks.The x-ray was used to verify a hypotube perforation/hole.The complaint was confirmed for under-pressure issues related to a hypotube perforation/hole.
 
Event Description
It was reported that tip of the catheter was hydroped and ruptured occurred.The target lesion was located in the lower extremity vein.An angiojet solent omni was selected for use for thrombectomy procedure.During the procedure, error was reported, and the tip of the balloon was hydroped and ruptured.The procedure was successfully completed with another of the same device and the patient was stable.It was further reported that an error message occurred, pump assembly leakage was noted, and shaft rupture occurred 5cm from the hub.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET SOLENT OMNI
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18722153
MDR Text Key336419006
Report Number2124215-2024-08730
Device Sequence Number1
Product Code DXE
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number45031
Device Catalogue Number45031
Device Lot Number0031436745
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2024
Initial Date FDA Received02/16/2024
Supplement Dates Manufacturer Received02/20/2024
Supplement Dates FDA Received03/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexFemale
-
-