Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
Philips received a complaint from the customer, reporting a machine and proximal pressure sensors failed message occurred on the v60 ventilator.The device was not in clinical use.There was no report of harm or other patient / user impact.
 
Manufacturer Narrative
H10: a manufacturer's product support engineer (pse) evaluated the device and confirmed diagnostic code 1007 (machine and proximal pressure sensors failed) indicated a problem with the data acquisition (da) printed circuit board assembly (pcba).The pse inspected the device further to isolate the da pcba failure and found the pcba data cable was disconnected.The cable was reconnected, and the device was left running to see if the fault persisted.There was no recurrence, and a complete performance verification testing (pvt) was completed.The client's configuration was maintained.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18722487
MDR Text Key336185870
Report Number2518422-2024-07986
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053614
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/09/2024
Initial Date FDA Received02/16/2024
Supplement Dates Manufacturer Received03/01/2024
Supplement Dates FDA Received03/04/2024
Date Device Manufactured10/03/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-