• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2024
Event Type  malfunction  
Event Description
Philips received a complaint by the customer on the v60, indicating that the device was getting error code 1122 "check vent: blower temperature high message." the device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.The device was swapped out with a different device.No medical intervention provided to the patient, nor a delay was noted.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The biomed customer was reporting running the unit for few days but is unable to reproduce the issue.The customer was reporting the error has been logged multiple times.Diagnostic code 1122- blower temperature high.The blower temperature was greater than 95ºc for longer than 60 seconds.The rse provided steps for additional troubleshooting.1.Verify that the air inlet filter is not dirty or blocked.2.Verify that the cooling fan is operational.3.Replace the blower.4.Replace the mc pcba.The rse provided the customer parts id for the v60 blower.
 
Manufacturer Narrative
H10: multiple attempts have been made on 16feb2024, 22feb2024, and 27feb2024 to try to obtain further information about this case, but no response received from the customer.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18725217
MDR Text Key335611383
Report Number2518422-2024-08050
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received02/16/2024
Supplement Dates Manufacturer Received03/08/2024
Supplement Dates FDA Received03/08/2024
Date Device Manufactured07/29/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-